Validation Protocol for Temperature & Humidity Monitoring Solutions (TMS) (IQ / OQ / PQ) – WHO-GMP Compliant

  1. Purpose

To establish documented evidence that the Temperature & Humidity Monitoring Solutions (TMS) consistently performs as intended and complies with WHO-GMP requirements through Installation Qualification (IQ), Operational Qualification (OQ), and Performance Qualification (PQ).

  1. Scope

This protocol applies to all temperature and humidity data loggers, sensors, software, storage devices, alarms, and reporting systems used in:

  • Warehouses
  • Cold rooms / refrigerators
  • Production areas
  • Stability chambers
  • Finished goods storage
  1. Responsibilities
RoleResponsibility
QAApproval, review, deviation handling
EngineeringInstallation and calibration
Production / StoresSystem usage
IT / AutomationSoftware & data integrity
ValidationExecution and documentation
  1. System Description
  • Temperature sensor (± accuracy as per spec)
  • Data logger with RTC (date & time stamp)
  • Storage (SD card / EEPROM / server)
  • Software for data analysis & report generation
  • Alarm mechanism (visual / audible / email / SMS)
  1. Installation Qualification (IQ)

Objective

To verify that the TMS is installed as per approved specifications.

IQ Checks

  • Equipment installed as per layout & drawings
  • Sensor placement at identified critical points
  • Power supply and backup verified
  • Firmware/software version recorded
  • Calibration certificates available & traceable
  • SOPs, manuals, and wiring diagrams available

Acceptance Criteria

✔ All components installed correctly
✔ Calibration certificates valid
✔ No installation deviations

  1. Operational Qualification (OQ)

Objective

To verify that the system operates within defined limits.

OQ Tests

  • Temperature measurement accuracy test (low, mid, high range)
  • RTC date/time accuracy verification
  • Data logging interval verification
  • Alarm activation at upper & lower limits
  • Power failure & data retention test
  • User access & data protection test

Acceptance Criteria

✔ Readings within tolerance
✔ Alarms triggered correctly
✔ No data loss during power interruption

  1. Performance Qualification (PQ)

Objective

To confirm consistent performance under actual operating conditions.

PQ Tests

  • Continuous monitoring for 7–30 days
  • Trend analysis for stability
  • Excursion simulation & response verification
  • Data retrieval, report generation & review

Acceptance Criteria

✔ Continuous accurate data
✔ Excursions detected & logged
✔ Reports acceptable for audits

  1. Deviations & CAPA

All deviations shall be documented, investigated, and closed with appropriate Corrective and Preventive Actions (CAPA).

  1. Validation Conclusion

Upon successful completion of IQ, OQ, and PQ, the TMS is deemed validated and GMP compliant.